System-related commands, Display {off | on, Display – Agilent Technologies E3632A User Manual

Page 95: Display:text <quoted string, Display:text, Display:text:clear, Output {off | on, Output, System-related commands 92

Advertising
background image

Chapter 4 Remote Interface Reference

System-Related Commands

92

System-Related Commands

DISPlay {OFF | ON}

This command turns the front-panel display off or on. When the display is

turned off, outputs are not sent to the display and all annunciators are disabled

except the

ERROR

annunciator.

The display state is automatically turned on when you return to the local mode.

Press the

key to return to the local state from the remote interface.

DISPlay?

This command queries the front-panel display setting. Returns ‘‘0’’ (OFF) or

‘‘1’’ (ON).

DISPlay:TEXT <quoted string>

This command displays a message on the front panel. The power supply will

display up to 12 characters in a message; any additional characters are

truncated. Commas, periods, and semicolons share a display space with the

preceding character, and are not considered individual characters.

DISPlay:TEXT?

This command queries the message sent to the front panel and returns a quoted

string.

DISPlay:TEXT:CLEar

This command clears the message displayed on the front panel.

OUTPut {OFF | ON}

This command enables or disables the outputs of the power supply. When the

output is disabled, the voltage value is 0 V and the current value is 20 mA. At

*RST

, the output state is OFF.

OUTPut?

This command queries the output state of the power supply. The returned value

is ‘‘0’’ (OFF) or ‘‘1’’ (ON).

Local

Advertising