Chapter 2 Initial Operation

Output Checkout

31

2

6

Ensure that the current can be adjusted from zero to the full rated

value.
Adjust the knob until the ammeter indicates 0 amps and then until the ammeter

indicates 7.0 amps.

1

N o t e

If an error has been detected during the output checkout procedures, the

ERROR

annunciator will turn on. See "Error Messages" for more information

starting on page 121 in chapter 5.

1

You can use the resolution selection keys to move the blinking digit the right

or left when setting the current.