Chapter 3 Front-Panel Operation

System-Related Operations

53

3

Error Conditions

When the front-panel

ERROR

annunciator turns on, one or more command

syntax or hardware errors have been detected. A record of up to 20 errors can

be stored in the power supply’s error queue.

See chapter 5 “Error Messages”,

starting on page 121 for a complete listing of the errors.

• Errors are retrieved in first-in-first-out (FIFO) order. The first error returned

is the first error that was stored. When you have read all errors from the queue,

the

ERROR

annunciator turns off. The power supply beeps once each time an

error is generated.

• If more than 20 errors have occurred when you operate the power supply over

the remote interface, the last error stored in the queue (the most recent error)

is replaced with -350, “

Too many errors

”. No additional errors are stored until

you remove errors from the queue. If no errors have occurred when you read

the error queue, the power supply responds with +0, “

No error

” over the remote

interface or “NO ERRORS” from the front panel.

• The error queue is cleared when power has been off or after a

*CLS

(clear

status) command has been executed. The

*RST

(reset) command does not

clear the error queue.

• Front-panel operation:

If the

ERROR

annunciator is on, press the

key repeatedly to read the

errors stored in the queue. All errors are cleared when you read all errors.

• Remote interface operation:

SYST:ERR?

Reads and error from the error queue

Errors have the following format (the error string may contain up to 80

characters).

-113, "Undefined header"

ERR

-113

Error