Disabling the Output Using an External Relay

51

3

Disabling the Output Using an External Relay

When the output of the E3632A is turned off, it is implemented by setting the

output to 0 volts and 0.02 amps. This gives a zero output voltage without

actually disconnecting the output. To disconnect the output an external relay

must be connected between the output and the load. A TTL signal of either low

true or high true is provided to control an external relay. This signal can only

be controlled with the remote command

OUTPut:RELay {OFF|ON}

. The TTL

output is available on the RS-232 connection pin 1 and pin 0.

When the

OUTPut:RELay

state is “ON”, the TTL output of pin 1 is high

(4.5 V) and pin 9 is low (0.5 V). The levels are reversed when the

OUTPut:RELay

state is “OFF”.

N o t e

TTL output of pin 1 or pin 9 of the RS-232 connector is available only after

installing two jumpers inside the power supply. See the Service Guide for

more information.

N o t e

Do not use the RS-232 interface if you have configured the power supply to

output relay control signals. Internal components on the RS-232 circuitry

may be damaged.

Knob Locking

The knob locking function can be used to disable the knob, thereby preventing

any unwanted changes during an experiment, or when you leave the power

supply unattended. To disable the knob, press the resolution selection key until

the blinking digit disappears.

Notice that the knob and front panel keys are disabled when in the remote

interface mode.