1 turn on the power supply, 3 enable the output, 4 adjust the voltage limit value to 1.0 volt – Agilent Technologies E3632A User Manual

Page 32: Current output checkout 30, Current output checkout

Advertising
background image

Chapter 2 Initial Operation

Output Checkout

30

Current Output Checkout

The following steps check basic current functions with a short across the

power supply’s output.

1

Turn on the power supply.

The power supply will go into the

power-on / reset

state; the output is disabled

(the

OFF

annunciator turns on); the 15V/7A range is selected (the

15V

annunciator turns on); and the knob is selected for

voltage

control.

2

Connect a short across (+) and (-) output terminals with an insulated

test lead.

3

Enable the output.
The

OFF

annunciator turns off and the

15V

,

OVP

, and

OCP

annunciators are lit.

The

CV

or

CC

annunciator turns on depending on the resistance of the test

lead. The

blinking

digit can be adjusted by turning the knob. Notice that the

display is in the meter mode. ‘‘Meter mode’’ means that the display shows the

actual output voltage and current.

4

Adjust the voltage limit value to 1.0 volt.
Set the display to the

limit

mode (the

Limit

annunciator will be

blinking

).

Adjust the voltage limit to 1.0 volt to assure CC operation. The

CC

annunciator

will turn on.

5

Check that the front-panel ammeter properly responds to knob control

for the 15V/7A range.
Set the knob to the

current

control, and turn the knob clockwise or counter

clockwise when the display is in the meter mode (the

Limit

annunciator is off).

Check that the ammeter responds to knob control and the voltmeter indicates

nearly zero (the voltmeter will show the voltage drop caused by the test lead).

Power

Output On/Off

Display Limit

Volt/Curr