Chapter 3 Front-Panel Operation

Programming Overvoltage Protection

44

3

Clear the overvoltage condition and exit this menu.
Now, when you press

key again, the “DONE” message is

displayed for a second and the

OVP

annunciator will not blink any more. The

output will return to meter mode.

• Remote interface operation:

VOLT:PROT {<voltage>|MIN|MAX}

Set the OVP level

VOLT:PROT:STAT {OFF|ON)

Disable or enable the OVP circuit

VOLT:PROT:CLE

Clear the tripped OVP circuit

N o t e

The power supply’s OVP circuit contains a crowbar SCR, which effectively

shorts the output of the power supply whenever the overvoltage condition

occurs. If external voltage source such as a battery is connected across the

output, and the overvoltage condition inadvertently occurs, the SCR will

continuously sink a large current from the source; possibly damaging the

power supply. To avoid this a diode must be connected in series with the

output as shown below.

Figure 3-1. Recommended Protection Circuit for Battery Charging

Over Voltage

Over Voltage