Chapter 3 Front-Panel Operation

Front-Panel Operation Overview

35

3

Front-Panel Operation Overview

The following section describes an overview of the front-panel keys before

operating your power supply.

• The power supply is shipped from the factory configured in the

front-panel

operation mode. At power-on, the power supply is automatically set to

operate in the front-panel operation mode. When in this mode, the front-

panel keys can be used. When the power supply is in

remote

operation mode,

you can return to front-panel operation mode at any time by pressing the

key if you did not previously send the front-panel lockout command.

A change between front-panel and remote operation modes will

not

result

in a change in the output parameters.

• The power supply has two output ranges of 15V/7A or 30V/4A. This feature

allows more voltage at a lower current or more current at a lower voltage.

The desired output range is selected from the front panel or over the remote

interfaces. The

15V

or

30V

annunciator indicates the presently selected

range.

• When you press

key (the

Limit

annunciator blinks), the

display of the power supply goes to the

limit

mode and the present limit

values will be displayed. In this mode, you can also observe the change of

the limit values when adjusting the knob. If you press the

key again or let the display time-out after several seconds, the power supply

will return the display to the

meter

mode (the

Limit

annunciator turns off).

In this mode, the actual output voltage and current will be displayed.

• The output of the power supply can be enabled or disabled from the front

panel using the

key. When the output is off, the

OFF

annunciator turns on and the output is disabled.

• The display provides the present operating status of the power supply with

annunciators and also informs the user of error codes. For example, the

power supply is operating in CV mode in the 15V/7A range and controlled

from the front panel, then the

CV

and

15V

annunciators will turn on. If,

however, the power supply is remotely controlled, the

Rmt

annunciator will

also turn on, and when the power supply is being addressed over GPIB

interface, the

Adrs

annunciator will turn on. See "

Display Annunciators

"

on page 5 for more information.

Local

Display Limit

Display Limit

Output On/Off