Chapter 3 Front-Panel Operation

Calibration Overview

69

3

To Change the Security Code

To change the security code, you must first

unsecure the power supply, and then enter a new code.

Be sure to read the security code rules on page 66 before attempting to secure

the power supply.

• Front-Panel Operation:

To change the security code, first make sure that the power supply is

unsecured

. Press

key after the “CAL MODE” message is displayed in

the calibration mode, enter the new security code using the knob and

resolution selection keys, then press

key.

Changing the code from the front panel also changes the code required from

the remote interface.

• Remote Interface Operation:

CAL:SEC:CODE <new code>Change the security code

To change the security code, first unsecure the power supply using the old

security code. Then, enter the new code. For example,

"CAL:SEC:STAT OFF, HP003632*"

Unsecure with old code

"CAL:SEC:CODE ZZ001443"

Enter new code

"CAL:SEC:STAT ON, ZZ00143"

Secure with new code