4 input 1 set up wiring, For your particular input (subsection 5.4 or – Honeywell UDC2500 User Manual

Page 154

Advertising
background image

Input Calibration

140

UDC2500 Universal Digital Controller Product Manual

4/07

Type of Input

Equipment Needed

RTD (Resistance
Thermometer
Device)

• A decade box, with at least ± 0.02 % accuracy, capable of providing

stepped resistance values over a minimum range of 0 to 1650 ohms
with a resolution of 0.001 ohm.

• Three insulated copper leads of equal length for connecting the

decade box to the controller.

Milliampere,
Millivolt, Volts, and
Radiamatic

• A calibrating device with at least ± 0.02 % accuracy for use as a

signal source.

• Two insulated copper leads for connecting the calibrator to the

controller.

• Place current source at zero before switching ON.
• Do not switch current sources OFF/ON while connected to the

UDC2500 input.

5.4 Input 1 Set Up Wiring

Thermocouple Inputs Using an Ice Bath

Refer to Figure 5-2 and wire the controller according to the procedure given in Table
5-4..

Table 5-4 Set Up Wiring Procedure for Thermocouple Inputs

Using an Ice Bath

Step

Action

1

Connect the copper leads to the calibrator.

2

Connect a length of thermocouple extension wire to the end of each copper lead and
insert the junction points into the ice bath.

3

Connect the thermocouple extension wires to the terminals for Input #1.
See Figure 5-2.

+
_

Millivolt

Source

26
27

+

Ice Bath

Copper Leads

Thermocouple
Extension Wire

_

+
_

Millivolt

Source

26
27

+

Ice Bath

Copper Leads

Thermocouple
Extension Wire

_

Figure 5-2 Wiring Connections for Thermocouple Inputs Using an Ice Bath

Advertising