Xylem 6-Series Multiparameter User Manual

Page 112

Advertising
background image

Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-106


NOTE: The number of items on this menu depends greatly on the sensors that are available and enabled on

your sonde. Below we describe every possible item on this menu. Your sonde probably may not have

every item described below.

To edit one of the following menu items choose the number or letter that corresponds to it.

TDS constant=0.65 This selection allows you to set the constant used to calculate TDS. TDS in g/L is

calculated by multiplying this constant times the specific conductance in mS/cm. This

item will only appear if the conductivity sensor is enabled in the “Sensors enabled”

menu. See Section 53, Principles of Operation for more information on the TDS

parameter.

Salinity=0

This selection allows you to input a manually-acquired value of salinity for calculating

other parameters such as DO mg/L and depth. This item is not used or displayed if the
conductivity sensor is enabled in the Sensors menu.

Pres=0 psi

This selection allows you to set a value of pressure for calculating other parameters like

salinity. This item is not used or displayed if the pressure sensor is enabled in the
Sensors menu.

Latitude=40 This selection allows you to input the global position (latitude) where the sonde is

sampling. The units are degrees. For accuracy, enter the decimal equivalent to indicate

degrees and minutes. For example, enter 41

o

30 ' as 41.5. This value is used in the

calculation of depth or level to account for global variations in the gravitational field.
This item will only appear if the pressure sensor is enabled in the Sensors menu.

Flow Setup… This selection allows you to setup the sonde to output flow information. See Appendix F,

Flow for detailed information on how to setup the Flow parameter. This item will only
appear if the sonde has shallow vented level.

Altitude=0

This selection allows you to input the local altitude (relative to sea level) where the sonde

is sampling. The units are feet. You may enter positive or negative values (range is -276

to 29028) to represent altitudes above or below sea level. This value is used in the

------------Advanced sensor-----------
1-TDS constant=0.65
2-Latitude=40
3-Altitude Ft=0
4-(*)Fixed probe
5-( )Moving probe
6-DO temp co %/C=1.1
7-DO warm up sec=40
8-(*)Wait for DO
9-Wipes=1
A-Wipe int=1
B-SDI12-M/wipe=1
C-Turb temp co %/C=0.3
D-(*)Turb spike filter
E-Chl temp co %/C=0


Select option (0 for previous menu):

Advertising
This manual is related to the following products: