Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-17

2.3.3

STEP 3 - POWER

Some type of external power supply is required to power the YSI 600R, 600QS, 600XL 6820V2-1, and the

non-battery version of the 600 OMS V2-1sondes. The YSI 6920V2-1, 6920V2-2, 6600V2-2,

6600EDS V2-2, 6600V2-4, 600XLM, and battery version of the 600 OMS V2-1sondes have internal

batteries or can run on external power.

If you have purchased a YSI 650 MDS display/logger, attaching your sonde to the display/logger will allow

your sonde to be powered from the batteries or the external power of the display/logger. See Section 3,

Displays/Loggers, for power options.

The battery-powered version of this instrument is powered by alkaline batteries, which the user must

remove and dispose of when the batteries no longer power the instrument. Disposal requirements vary by

country and region, and users are expected to understand and follow the battery disposal requirements for

their specific locale.

The circuit board in this instrument contains a manganese dioxide lithium "coin cell" battery that must be in

place for continuity of power to memory devices on the board. This battery is not user serviceable or

replaceable. When appropriate, an authorized YSI service center will remove this battery and properly

dispose of it, per service and repair policies.

POWER FOR LAB CALIBRATION

A YSI 6038 (110 VAC) or 6651 (64-240 VAC) Power Supply is required for sondes without internal

batteries when using them with a PC for calibration and setup. Sondes with internal batteries do not require

a power supply, but using the sonde with a power supply in the lab is often convenient and extends battery

life. Most adapters include a short pigtail for power that plugs into the power supply. After attaching the

four-pin connector from the power supply to the pigtail, simply plug the power supply into the appropriate

AC outlet.

See Section 2.2, Connecting Your Sonde, for specific information on cables, adapters and power supplies

required for connecting your sonde to various devices.

Figure 32

The system configuration best suited for

initial setup is shown in Figure 32.