Calibrate TSS

TSS is an abbreviation for Total Suspended Solids. TSS is a unit derived from a

Turbidity (NTU) measurement and correlated TSS-NTU measurements. The Calibrate

TSS feature is only used if the TSS parameter has been added to the current data file

using Add/Remove Parameters. This command opens the TSS Calibration window and

allows calibration adjustment.

How do I add TSS to my data file?

You must have a data set that includes Turbidity data points.

It is also common to use a 0 to 0 correlation as a base point in these measurements. This

can be used as the first correlation point. You need to establish a correlation between

your Turbidity reading and a TSS value. This is done by taking a sample and noting its

Turbidity in NTU’s. Then run a lab analysis to obtain a total suspended solids reading.

You have now determined one correlation point.