Xylem 6-Series Multiparameter User Manual

Page 12

Advertising
background image

Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-6

DO MEMBRANE INSTALLATION WITH THE PROBE NOT INSTALLED IN THE

SONDE

Remove the protective cap and the dry membrane from the YSI 6562 Dissolved Oxygen probe.

Make sure that the protective cap is installed on the connector end of the probe. Do not allow the

electrolyte solution to wet the probe‟s connector and O-ring seal areas. This solution is extremely corrosive

to the connector and is difficult to remove.

Before using any electrolyte, wrap a clean, dry paper towel around the capped probe to catch any spilled

electrolyte. Hold the probe in a vertical position and apply a few drops of KCl solution to the tip. The fluid

should completely fill the small moat around the electrodes and form a meniscus on the tip of the sensor.

Be sure no air bubbles are stuck to the face of the sensor. If necessary, shake off the electrolyte and start

over.

Figure 7

Secure a membrane between your left thumb and the probe body.

Always handle the membrane with care, touching it only at the

ends.

With the thumb and forefinger of your right hand, grasp the free

end of the membrane. With one continuous motion, gently stretch

it up, over, and down the other side of the sensor. The membrane

should conform to the face of the sensor.

Secure the end of the membrane under the forefinger of your left

hand.

Roll the O-ring over the end of the probe, being careful not to

touch the membrane surface with your fingers. There should be no

wrinkles or trapped air bubbles. Small wrinkles may be removed

by lightly tugging on the edges of the membrane. If bubbles are

present, remove the membrane and repeat steps 3-8.

Trim off any excess membrane with a sharp knife or scissors.

Rinse off any excess KCl solution, but be careful not to get any

water in the connector.

If you are concerned that electrolyte may have dripped onto the O-

ring seal area, probe connector, or bulkhead connectors, rub the

area clean with paper towels wetted with Deionized water and then

dry the affected area with a final dry towel, compressed air blasts,

or rinse with fresh alcohol.

NOTE: You may find it more convenient to mount the probe vertically in a vise with rubber jaws while

applying the electrolyte and membrane to the sensor tip.

3

4

5

6

7

8