= 0.6). Even if the ionic content were due to sodium chloride, the ammonium reading would be

erroneously high, about 0.4 mg/L.


However, brackish or seawater has enough sodium and potassium to cause interference so great as to
make the sensor unsuitable for these media.

The sensor used in the sonde detects only ammonium ions (NH

4

+

), the predominant form of total

ammonium nitrogen in most environmental samples. However, using the concurrently determined values

of pH, temperature, and conductivity, the sonde software can also provide the user with the concentration

of free ammonia (NH

3

) in the sample under investigation.

Ammonium ions and free ammonia are in equilibrium in any solution according to the following equation:

NH

4

+

NH

3

+ H

+

The value of the equilibrium constant associated with this reaction, K = [NH

3

][H

+

]/[NH

4

+

], and its variation

with temperature and salinity, is well known. This information allows the free ammonia concentration

[NH

3

] to be automatically calculated by the sonde software and displayed if this parameter is activated.

Despite the potential problems with interference when using ISEs, it is important to remember that almost

all interfering species produce an artificially high ammonium reading. Thus, if the sonde indicates the

presence of only small quantities of ammonium, it is unlikely that the reading is erroneously low because of

interference. Unusually high ammonium readings (which could be due to interfering ions) should be

confirmed by laboratory analysis after collection of water samples.

Of all the sensors available on the sonde, ion selective electrodes have the greatest tendency to exhibit

calibration drift over time. This drift should not be a major problem for sampling studies where the

instrument can be frequently calibrated. However, if an ammonium sensor is used in a longer-term

deployment study with the sonde, the user should be aware that drift is almost certain to occur. The extent

of the drift will vary depending on the age of the probe, the flow rate at the site, and the quality of the

water. For all monitoring studies using ion selective electrodes, the user should acquire a few “grab

samples” during the course of the deployment for analysis in the laboratory by chemical means or with

another ammonium sensor which has been recently calibrated. Remember that the typical accuracy

specification for the sensor (+/- 10 % of the reading or 2 mg/L, whichever is larger) refers to sampling

applications where only minimal time has elapsed between calibration and field use.

CALIBRATION AND EFFECT OF TEMPERATURE

The ammonium sensor should be calibrated using solutions of known total ammonium-nitrogen content

according to the procedures detailed in Sections 2.6.1 and 2.9.2. If a two point calibration protocol is used,

the temperature of the standards should be as close as possible to that of the environmental medium to be

monitored. The recommended calibration procedure is one involving three solutions. Two of the solutions

should be at ambient temperature while the third should be at least 10 degrees Celsius different from

ambient temperature. This protocol minimizes the effects of taking readings at temperatures that are

significantly different ambient laboratory temperatures.

MEASUREMENT AND CALIBRATION PRECAUTIONS

(1) The temperature response of ion selective electrodes is not as predictable as that of pH sensors.

Therefore, be sure to carry out a 3-point calibration the first time you use the probe. This will provide a

default setting for the effect of temperature on your particular sensor. After this initial calibration, you can

use the less time consuming 2 point and 1-point routines to update the 3-point calibration. However, we

strongly recommend a new 3-point calibration after each deployment of 30 days or longer.