Select 1 – Start sampling option to start discrete sampling. After the initial sampling time interval has

passed (4 seconds in the screen above), sequential lines of data will appear on the screen.

The following prompt will appear just below the screen header:

1-LOG last sample, 2-LOG ON/OFF, 3-Clean optics

By entering 1–LOG last sample, A single line of data can be logged to sonde memory and the

following message will be displayed: Sample logged.

By entering 2-LOG ON/OFF, a set of data can be logged to memory and the following message will

be displayed: LOG is ON, hit 2 to turn it OFF, 3-Clean optics. Press 2 again to terminate logging.

By entering 3-Clean optics, if your unit has an optical probe, the wiper will clean the optical surface.

The 3-Clean optics portion of the prompt will only appear if an optical probe was installed and

enabled.

Select Esc or press 0 and return to the Discrete Sampling menu.

From the Discrete Sampling menu, select 2-Sample Interval to type a number that represents the number

of seconds between samples. The maximum sample interval is 32767 seconds (9+ hours). The factory

default sample interval is 4 seconds and works best for most discrete sampling applications. The shortest

possible sample interval for obtaining new DO readings is 4 seconds. If a smaller interval is selected, then

the DO readings remain constant for the number of samples necessary to fill 4 seconds.

NOTE: If your have used your sonde with a 650 MDS data logger, the Sample Interval automatically be

changed to 0.5 seconds.

=============================================================================

Date Time Temp SpCond Sal DO Depth pH Turbid Battery

mm/dd/yy hh:mm:ss C mS/cm ppt mg/L feet NTU volts
------------------------------------------------------------------------------
*** 1-LOG last sample 2-LOG ON/OFF, 3-Clean optics***
05/05/97 10:04:40 21.57 0.009 0.00 7.73 -0.293 7.55

0.3 10.2

05/05/97 10:04:44 21.57 0.009 0.00 8.41 -0.300 7.55

0.3 10.3

*** LOG is ON, hit 2 to turn it OFF, 3-Clean optics***
05/05/97 10:04:48 21.58 0.009 0.00 8.41 -0.302 7.55

0.3 10.3

05/05/97 10:04:52 21.58 0.009 0.00 8.40 -0.302 7.55

0.3 10.3

05/05/97 10:04:56 21.58 0.009 0.00 8.41 -0.303 7.55

0.3 10.2

05/05/97 10:05:00 21.58 0.009 0.00 8.42 -0.303 7.55

0.3 10.3

*** 1-LOG last sample 2-LOG ON/OFF , 3-Clean optics***
05/05/97 10:05:04 21.58 0.009 0.00 8.44 -0.304 7.55

0.3 10.3

Sample logged.
05/05/97 10:05:08 21.58 0.009 0.00 8.45 -0.305 7.55

0.3 10.3