Xylem 6-Series Multiparameter User Manual

Page 91

Advertising
background image

Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-85

When selecting Optic X -6026-Turbidity (or 6136-Turbidity), there will be a choice of 1-point, 2-point,

or 3-point calibrations for your turbidity sensor.

The 1-point option is normally used to zero the turbidity probe in 0 NTU standard. Place the sonde in clear

water (deionized or distilled) with no suspended solids, and input 0 NTU at the screen prompt. Press Enter

and the screen will display real-time readings that will allow you to determine when the turbidity readings

have stabilized. Press Enter after the readings have stabilized to confirm the calibration and zero the

sensor. Then, as instructed, press any key to return to the Calibrate menu.

Select the 2-point option to calibrate the turbidity probe using only two calibration standards. In this case,

one of the standards must be clear water (0 NTU) and the other should be in the range of known turbidity

for the water to be monitored. For example, if the water to be evaluated is known to be low in turbidity, an

appropriate choice of standards might be 0 and 10 NTU (6026) or 0 and 12.7 NTU (6136). However, for

general purpose measurements an appropriate choice of standards is usually 0 and 100 NTU (6026) or 0

and 126 NTU (6136).

To begin the calibration, immerse the sonde in the 0 NTU standard, as instructed, and press Enter. It is

mandatory that the 0 NTU standard be calibrated first. The screen will display real-time readings that will

allow you to determine when the readings have stabilized. Pressing Enter will confirm the first calibration.

Following the instructions on the screen, place the sonde in the second turbidity standard, input the correct

turbidity value in NTU, press Enter, and view the stabilization of the values on the screen in real-time.

After the readings have stabilized, press Enter to confirm the calibration. Then, as instructed, press any

key to return to the Calibrate menu.

Select the 3-point option for maximum accuracy over the entire range of 0 to 1000 NTU. As for the 2-

point procedure, one of the standards must be 0 NTU. Because of the linearity characteristics of the

sensors, we recommend that the other two standards have turbidity values of 100 and 1000 NTU.

However, the user can select any values that are deemed appropriate. The procedure for this calibration is

the same as for a 2-point calibration, but the software will prompt you to place the sonde in the additional

solution to complete the 3-point procedure.

For all turbidity calibration procedures, be certain that the standard and sensor are thermally equilibrated

prior to proceeding with the calibration.

For further information related to setting up, calibrating and running turbidity measurements, refer to

Appendix E, Turbidity Measurements.



OPTIC CHLOROPHYLL

When selecting Optic X - Chlorophyll, there will be a choice of zeroing the relative fluorescence unit

parameter (Chl RFU Zero) or calibrating with actual chlorophyll standards (µg/L 1-point, 2-point, or 3-

point).

If you have selected Chl RFU as a parameter in the Report menu, the sonde will report only relative

values of fluorescence in the sample being measured. These values could then be converted into actual

chlorophyll concentrations in µg/L by using a post-calibration procedure, after the chlorophyll content of

grab-samples taken during a sampling or monitoring study has been analyzed in a laboratory. This

determination can involve conducting the extractive analysis procedure described for chlorophyll in
Methods for the Examination of Water and Wastewater

or by carrying out an in situ measurement of

chlorophyll using a commercial benchtop fluorometer.

The Chl RFU Zero option is used to zero the fluorescence probe in a medium that is chlorophyll-free.

Place the sonde in clear water, and input 0 at the screen prompt. Press Enter and the screen will display

real-time readings that will allow you to determine when the fluorescence readings have stabilized. Press

Advertising
This manual is related to the following products: