Section 4

YSI Incorporated

Environmental Monitoring Systems Manual

4-37

Once you have at least two correlation points, you have established a linear correlation

between the two measurements. EcoWatch allows you to enter up to five correlation

points. The more correlation points you have, the higher that your accuracy will be.

To add the TSS parameter to your data file, go to the Setup menu. Choose Parameters,

then Add/Remove. Select the TSS parameter and click the Add button. The Calibrate

TSS window will appear.

Name the file for identification in the future. Multiple calibration files may be created.

Any previously created calibration files will be located in the drop-down name list box.

Once you have entered the desired number of correlation points, click OK. EcoWatch

will calculate the TSS values by doing a linear interpolation of the Turbidity (NTU) data

and the correlation chart .

GRAPH

The Setup menu, Graph submenu has the following commands:

Font/Color

Goes to the Font Dialog Box where you can change the font and color of text that appears in the

graph.

Font

Type or select a font name. EcoWatch lists the fonts available with the current

printer driver and additional fonts installed in your system.

Font Style

Select a font style. To use the default type style for a given font, select Regular.

Size

Type or select a font size. The sizes available depend on the printer and the selected

font. If the size you type is not available on the current printer, EcoWatch chooses

the closest available size.

Effects

Choosing Strikeout will draw a line through all text in the table. Choosing

underline will underline the text.