Xylem 6-Series Multiparameter User Manual

Page 58

Advertising
background image

Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-52

Skip B-View params to log in this initial test study. This feature will be explained in detail in Section

2.9, Sonde Menu.

After making the above entries, the sonde software will automatically calculate the expected battery life,

and the time it will take for the sonde memory to be filled. This information is displayed on the screen for

your consideration as items 8, 9, and A. If the battery life or the free memory capacity will be exceeded

sooner than the duration, you may want to make some changes to the entries. For example, you can free up

memory in the sonde by uploading all existing data from the sonde memory to your PC and then deleting

them out of the sonde (see 3-File from Main Menu). You may want to change the batteries for longer

battery life. You can lengthen the sampling interval to extend both battery life and memory capacity.

Review the screen below, which now displays the entries made above and reflects your logging conditions.

Once you press C-Start logging, a screen will appear to request confirmation.

Select 1-Yes and the screen will change.

------------Unattended setup-----------
1-Interval=00:15:00
2-Start date=07/17/96
3-Start time=18:00:00
4-Duration days=365
5-File=clrlake3
6-Site=Clear Lake at Spillway
7-Bat volts: 11.6
8-Bat life 25.1 days
9-Free mem 41.3 days
A-1st sample in 4.10 minutes
B-View params to log
C-Start logging

-------------Start logging-------------
Are you sure?
1-Yes
2-No

Select option (0 for previous menu):

Advertising
This manual is related to the following products: