Calibration and effect of temperature, Optical fiber, Photodetector – Xylem 6-Series Multiparameter User Manual

Page 267: Light source, Green led, Optical filter

Advertising
background image

Principles of Operation

Section 5

YSI Incorporated

Environmental Monitoring Systems Manual

5-29

Field optical measurements are particularly susceptible to fouling, not only from long term build up of

biological and chemical debris, but also to shorter term formation of bubbles from outgassing of the

environmental water. These bubbles can generally be removed in short term sampling application by

simply agitating the sonde manually. However, for studies longer than a few hours where the user is not

present at the site, the quality of the data obtained with a Rhodamine WT sensor that has no capability of

mechanical cleaning may be compromised. Like all YSI optical probes described above, the 6130

Rhodamine WT probe is equipped with a mechanical wiper that makes it ideal for unattended applications.

The wiper can be activated in real-time during discrete sampling operations or will function automatically

during long term unattended sampling studies. The number of wiper movements and the frequency of the

cleaning cycle for the unattended mode can be set in the sonde software. Generally one movement is

sufficient for most environmental application, but in media with particularly heavy fouling, additional

cleaning cycles may be necessary.

CALIBRATION AND EFFECT OF TEMPERATURE

The sonde software offers the option of 1 point, 2 point, or 3 point calibrations procedures. For most

applications, a 2-point calibration at 0 and 100 ug/L is sufficient. However, for maximum accuracy over

the entire 0-200 ug/L range of the sensor, a 3-point calibration procedure can slightly enhance the accuracy

of the sensor. Note that YSI does not offer Rhodamine WT standards, but does suggest a vendor for user-

production of standards and, later in this section, provides instructions for preparing a solution which is 100

ug/L in Rhodamine WT.

While the effect of temperature on the Rhodamine sensor and electronics is small, the fluorescence of

Rhodamine WT changes significantly with temperature. The combination of the two factors is

automatically taken into account by the sonde software providing temperature compensated readings.


Optical Fiber

Photodetector

Light Source

Green LED

Optical Filter