12 chloride – Xylem 6-Series Multiparameter User Manual

Page 254

Advertising
background image

Principles of Operation

Section 5

YSI Incorporated

Environmental Monitoring Systems Manual

5-16

(2) Ion selective electrodes may not stabilize as rapidly as pH sensors. Be sure to allow plenty of time for

the readings to come to their final values during all calibration routines.

(3) Ion selective electrodes generally drift more than pH sensors. To check for this drift, place the sonde in

one of your standards at the end of each deployment.

(4) Ammonium standards are good growth media for a variety of organisms. This growth can significantly

reduce the nitrogen content of your standards, an effect that is particularly important for the 1 mg/L

solution. It is best to use new standards for each deployment, but if you decide to save your solutions for

reuse, we recommend refrigerated storage to minimize the growth of these organisms.

(5) Remember that the ammonium sensor will take longer to stabilize after exposure to buffers in a sonde

pH calibration. To accelerate this process, soak the sensor in 100 mg/L standard for a few minutes after

performing a pH calibration. In addition, be particularly careful that readings are stable during ammonium

calibration after exposure to buffers.


CAUTION

: The ammonium membrane module is for use only at depths less than 50 feet (15.2 meters).

Use of the probe at greater depths is likely to permanently damage the sensor.

5.12 CHLORIDE

The sonde chloride probe employs a solid state membrane attached to a conductive wire. When the probe

is immersed in water, a potential is established across the membrane that depends on the amount of

chloride in the medium. This potential is read relative to the reference electrode of the sonde pH probe. As

for all ISEs, there is a linear relationship between the logarithm of the chloride activity (or concentration in

dilute solution) and the observed voltage. The Nernst equation describes this relationship.

Under ideal conditions, the Nernst equation predicts a response of 59 mV for every 10-fold rise in chloride

activity at 25°C. However, in practice, empirical calibration of the electrode is necessary to establish the slope

of the response. Typical slopes are

45-55 mV per decade for YSI sensors. This slope value is determined by

calibration with two solutions of known chloride concentration (typically 10 mg/L and 1000 mg/L Cl

-

).

The slope of the plot of log (chloride) vs. voltage is also a function of temperature, changing from its value at

calibration by a factor of the ratio of the absolute temperatures at calibration to that at measurement. The

point where this new plot of log (chloride) vs. voltage intersects the calibration plot is called the isopotential

point, that is, the chloride concentration at which changes in temperature cause no change in voltage. Our

experience with ISEs indicates that for best accuracy, the isopotential point should be determined empirically.

To do so, the user employs a third calibration point where the voltage of the lower concentration standard is

determined at a temperature at least 10°C different from the first two calibration points. The slope, offset and

isopotential point drift slowly, and the probe should be recalibrated periodically.

All ion selective electrodes are subject to the interaction of species with the sensor membrane, which are

similar in nature to the analyte. These interfering species thus include other halide ions (fluoride, bromide,

and iodide) as well as other anions.

Despite the potential problems with interference when using ISEs, it is important to remember that almost

all interfering species produce an artificially high chloride reading. Thus, if the sonde indicates the presence

of only small quantities of chloride, it is unlikely that the reading is erroneously low because of

interference. Unusually high chloride readings (which could be due to interfering ions) should be

confirmed by laboratory analysis after collection of water samples.

Of all the sensors available on the sonde, ion selective electrodes have the greatest tendency to exhibit

calibration drift over time. This drift should not be a major problem for sampling studies where the

Advertising
This manual is related to the following products: