4 logging barometer readings – Xylem 6-Series Multiparameter User Manual

Page 186

Advertising
background image

650 MDS

Section 3

YSI Environmental Monitoring Systems Operations Manual

3-56

3.9.4 LOGGING BAROMETER READINGS

To log barometric pressure readings with the 650, the following conditions (similar to those above for

logging GPS readings) must be met:

Barometer readings can ONLY be logged to 650 memory – the system does NOT support logging to

sonde memory.

The barometer readings must be logged to 650 memory from the 650 Run display along with sonde

sensor readings. This means that you must have a sonde attached to your 650 in order to log barometer

readings. Logging of barometer readings is not supported without sonde attachment even though these

readings can be viewed in the Status Bar with no sonde attached.

If your want to log barometer readings to 650 memory, you must make certain that the Store

Barometer option is active in the Logging setup menu of the 650.

Once these conditions are met, follow the instructions in Section 3.5.3 above for logging of sonde sensor

readings to 650 memory. The barometer readings will automatically be added to the file at the sample interval

selected for the sonde readings.

Barometer readings stored in 650 memory will remain as part of the sensor data file that can be viewed

from the 650 display and/or uploaded with the data file for analysis in EcoWatch for Windows. The

barometric pressure readings are processed as a normal parameter by EcoWatch, with the user having the

ability to change units from the EcoWatch menus. For example, data stored in 650 memory in units of mm

Hg units can easily be converted to readings in mBar after upload.