3 sensor performance problems – Xylem 6-Series Multiparameter User Manual

Page 285

Advertising
background image

Troubleshooting

Section 6

YSI Incorporated

Environmental Monitoring Systems Operations Manual

6-

3

6.3 SENSOR PERFORMANCE PROBLEMS


SYMPTOMS

POSSIBLE CAUSE

ACTION

REF

Rapid Pulse dissolved
Oxygen reading
unstable or inaccurate

Probe not properly calibrated

Follow DO cal procedures

2.6.1

Membrane not properly installed or
may be punctured

Follow 6562 setup procedure

2.3.1

DO probe electrodes require
cleaning

Follow DO cleaning procedure.
Use 6035 maint. kit

2.10.2

Water in probe connector

Dry connector; reinstall probe

2.3.2

Algae or other contaminant
clinging to DO probe

Rinse DO probe with clean water

2.10.2

Barometric pressure entry is
incorrect

Repeat DO cal procedure

2.6.1

Cal at extreme temperature

Recal at (or near) sample
temperature

2.6.1

DO Charge too high ( >75)
1. Anodes polarized (tarnished)
2. Probe left on continuously

Enable DO charge parameter in
the Sonde report menu. Run
sonde, if charge is over 100,
recondition probe with 6035
Maintenance Kit. Follow DO
cleaning procedure.


2.10.2

DO Charge too low (<25)
Insufficient electrolyte.

Replace electrolyte and
membrane.

2.3.1

DO probe has been damaged

Replace 6562 probe

9

Internal failure

Return sonde for service

9

ROX dissolved Oxygen
reading unstable or
inaccurate

Probe not properly calibrated

Follow DO cal procedures

2.6.1

Membrane assembly not properly
installed or may be punctured

Follow instructions for proper
membrane assembly replacement

2.10.2

Water in probe connector

Dry connector; reinstall probe

2.3.2

Algae or other contaminant
clinging to DO probe

Rinse DO probe with clean water

2.10.2

Barometric pressure entry is
incorrect

Repeat DO cal procedure

2.6.1

Membrane assembly is more than
1 year old

Replace membrane assembly
(6155 Optical DO Membrane
Replacement Kit)

9

Wiper is not parking correctly

Make certain that setscrew is
tight. If problem persists, return
probe for service.

9

DO probe has been damaged

Replace 6150 probe

9

pH, ORP, chloride,
ammonium, or nitrate
readings are unstable or
inaccurate. Error
messages appear during
calibration.

Probe requires cleaning,

Follow probe cleaning procedure

2.10.2

Probe requires calibration

Follow cal procedures

2.6.1

pH probe reference junction has
dried out from improper storage.

Soak probe in tap water or buffer
until readings become stable

2.10.2