Specifications

Appendix O

YSI Incorporated

Environmental Monitoring Systems Operations Manual

O-3

Computer Interface:

RS-232C, SDI-12

Internal logging

memory size:

384 kilobytes (150,000 individual parameter readings)

Power:

8 C-size Alkaline Batteries or External 12 VDC

Battery Life:

Approximately 55 days at 20 C at a 15 minute logging interval, and temperature,

conductivity, pH/ORP, ROX Optical DO, Turbidity, Chlorophyll, and BGA-PC

(or BGA-PE) sensors active. Battery life is heavily dependent on sensor

configuration and is given above for a typical sensor ensemble. If you have a

different sensor configuration, set up your sonde for a deployment in the

Run|Unattended menu and check the projected approximate battery life.



6920V2-1 SONDE

Available Sensors

Temperature, Conductivity, Rapid Pulse Dissolved Oxygen, pH, ORP, three ion

selective electrodes (ammonium, nitrate, chloride), and Depth (shallow, medium,

shallow vented). One total optical sensor (ROX Optical DO, Turbidity,

Chlorophyll, Rhodamine WT, BGA-PC, or BGA-PE). Note that Rapid Pulse and

ROX DO sensors cannot be activated simultaneously.

Operating Environment Medium: fresh, sea, or polluted water

Temperature: -5 to +50

C for most sensors

Depth: 0 to 200 feet (61 meters)

Storage Temperature: -40 to +60

C for sonde and all sensors except pH , pH/ORP, ISE and optical

sensors

-10 to +60

C for pH, pH/ORP, ISE, and optical sensors

Material:

Polyurethane, PVC, Stainless Steel

Diameter:

2.9 inches (7.4 cm)

Length:

Approximately 18.25 inches (46.4 cm) with no depth.

Weight:

Approximately 3.74 pounds (1.7 kg).

Computer Interface:

RS-232C, SDI-12

Internal logging

memory size:

384 kilobytes (150,000 individual parameter readings)

Power:

8 AA-size Alkaline Batteries or External 12 VDC

Battery Life:

Approximately 30 days at 20

C at 15 minute logging interval with temperature,

conductivity, pH/ORP, and Rapid Pulse DO with a 40-second DO warm up time,

and one optical probe other than ROX optical DO active. Battery life is heavily

dependent on sensor configuration and is given above for a typical sensor

ensemble. If you have a different sensor configuration, set up your sonde for a

deployment in the Run|Unattended menu and check the projected approximate

battery life.