13 turbidity – Xylem 6-Series Multiparameter User Manual

Page 255

Advertising
background image

Principles of Operation

Section 5

YSI Incorporated

Environmental Monitoring Systems Manual

5-17

instrument can be frequently calibrated. However, if a chloride sensor is used in a longer-term deployment

study with the sonde, the user should be aware that drift is almost certain to occur. The extent of the drift

will vary depending on the age of the probe, the flow rate at the site, and the quality of the water. For all

monitoring studies using ion selective electrodes, the user should acquire a few “grab samples” during the

course of the deployment for analysis in the laboratory by chemical means or with another chloride sensor

which has been recently calibrated. Remember that the typical accuracy specification for the sensor (+/- 15

% of the reading or 5 mg/L, whichever is larger) refers to sampling applications where only minimal time

has elapsed between calibration and field use.

CALIBRATION AND EFFECT OF TEMPERATURE

The chloride sensor should be calibrated using solutions of known chloride content according to the

procedures detailed in Sections 2.6.1 and 2.9.2. If a two point calibration protocol is used, the temperature

of the standards should be as close as possible to that of the environmental medium to be monitored. The

recommended calibration procedure is one involving three solutions. Two of the solutions should be at

ambient temperature while the third should be at least 10 degrees Celsius different from ambient

temperature. This protocol minimizes the effects of taking readings at temperatures that are significantly

different ambient laboratory temperatures.

MEASUREMENT AND CALIBRATION PRECAUTIONS

(1) The temperature response of ion selective electrodes is not as predictable as that of pH sensors.

Therefore, be sure to carry out a 3-point calibration the first time you use the probe. This will provide a

default setting for the effect of temperature on your particular sensor. After this initial calibration, you can

use the less time-consuming 2 point and 1-point routines to update the 3-point calibration. However, we

strongly recommend a new 3-point calibration after each deployment of 30 days or longer.

(2) Ion selective electrodes may not stabilize as rapidly as pH sensors. Be sure to allow plenty of time for

the readings to come to their final values during all calibration routines.

(3) Ion selective electrodes generally drift more than pH sensors. Be sure to check for this drift by placing

the sonde in a standard at the end of each deployment.

(4) Remember that the chloride sensor will take longer to stabilize after exposure to pH buffers. To

accelerate this process, soak the sensor in 1000 mg/L standard for a few minutes after performing a pH

calibration. In addition, be particularly careful that readings are stable during chloride calibration after

exposure to buffers.

5.13 TURBIDITY

Turbidity is the measurement of the content of suspended solids (cloudiness) in water and is typically

determined by shining a light beam into the sample solution and then measuring the light that is scattered

off of the particles which are present. For turbidity systems capable of field deployment, the usual light

source is a light emitting diode (LED) which produces radiation in the near infrared region of the spectrum.

The detector is usually a photodiode of high sensitivity. The angle between the emitted and detected light

varies (usually between 90 and 180 degrees) depending on the probe used. The International Standards

Organization (ISO) recommends the use of a light source with a wavelength between 830 and 890 nm and

an angle of 90 degrees between the emitted and detected radiation (ISO 7027).

The turbidity system available as an option for use with YSI 6-series sondes consists of a probe which

conforms to the above ISO recommendations. The output of the sonde turbidity sensor is processed via the

Advertising
This manual is related to the following products: