Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-22

Check the O-rings and sealing surfaces for any contaminants that

could interfere with the seal of the battery chamber.


CAUTION:

Make sure that there are NO contaminants between

the O-ring and the sonde. Contaminants that are present under

the O-ring may cause the O-ring to leak when the sonde is

deployed.

Lightly lubricate the o-rings on the bottom of the threads and on the

connector stem as shown in Figure 37.

Return the battery lid and tighten by hand. DO NOT OVER-
TIGHTEN.

Figure 37



2.3.4 STEP 4 - CONNECTING A FIELD CABLE

Figure 38

All YSI 6600V2-2, 6600EDS V2-2, 6600V2-4,

6920V2-1, 6920V2-2, 6820V2-2, 600XLM, 600QS,

and 600 OMS V2-1sondes have a sonde-mounted

cable connector for attachment of the field cable.

Some versions of the 600R, 600XL, and 6820V2-1

sondes also have this connector.

However, some versions of the YSI 600R, 600XL,

and 6820V2-1sondes have permanently attached

“integral” cables. If your sonde has a cable that is

non-detachable, the next section will not be

relevant.

To attach a field cable to the sonde connector,

remove the waterproof cap from the sonde

connector and set it aside for later reassembly

during deployment or storage. Then connect your

field cable to the sonde connector.

A built-in “key” will ensure proper pin alignment. Rotate the cable gently until the “key” engages and then

tighten the connectors together by rotating clockwise. Attach the strain relief connector to the sonde bail.

Rotate the strain relief connector nut to close the connector's opening.

For all of the sondes, the other end of the cable is a military-style 8-pin connector (MS-8). Through use of

a YSI 6095B MS-8 to DB-9 adapter, the sonde may be connected to a computer for setup, calibration, real-

time measurement, and uploading files.

This MS-8 connector also plugs directly into the 650 MDS display/logger. This instrument contains a

microcomputer that allows it to be used in a similar manner to that of a terminal interface to a PC.

As an alternative to the field cable, you may use a YSI 6067B calibration cable for laboratory interaction

with the sonde. In this case, simply plug the proper end of the cable into the sonde connector and attach

the DB-9 connector of the cable to the Com port of your computer.

+

-

-

+

BATTERY CAP

BULKHEAD

CONNECTOR

BAIL

O-RINGS

REMOVE

WATERPROOF CAP

SONDE

CONNECTOR

FIELD CABLE

CONNECTOR

BAIL

STRAIN RELIEF

CONNECTOR