Xylem 6-Series Multiparameter User Manual

Page 125

Advertising
background image

Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-119

regeneration cycle has been successful. The felt filters should also be dried at about 100° C (about 200° F)

for about 30 minutes before assembly.

Desiccant material is sold separately. Both the cartridge and canister can easily be opened, emptied, and

refilled.

CAUTION

: It is important to keep the tube in vented sondes and cables dry. They are supplied with caps

for closing the volume when not in use. Keep the caps on until just before calibration and deployment. For

storage, replace the caps.


6882 CHLORIDE, 6883 NITRATE AND 6884 AMMONIUM SENSORS

You should clean these probes whenever deposits or contaminants appear on the ion selective membranes

located on the tips of these sensor modules. If possible, the

module should be cleaned while installed in the sonde

bulkhead. Use clean water and a moist piece of lens

cleaning tissue to gently wipe the sensor membranes until no

more contaminants are removed. However, under some

conditions, it may be necessary to remove the module from

the sonde bulkhead for cleaning and/or storage. To remove

the module, follow the diagram at the side. Remove the

module with finger pressure only, if possible. However, it is

acceptable to use small pliers if necessary to loosen the

module for final removal by hand. Be very careful not to

squeeze the module any more than is necessary for a firm

grip. Use slipjaw pliers to minimize the chance of

applying too much pressure on the module. While the

module is removed from the sonde, be sure to dry the sonde port and probe connector with compressed air

and apply a very thin coat of lubricant to the O-ring before re-installation.

NOTE: The ion selective membranes are fragile. Be certain to: (1) Use only moist, high quality lens

cleaning tissue for the cleaning procedure (not paper towels or other coarse materials); and (2) Stroke the

probe face very gently with the tissue during the procedure. If you do not adhere to these instructions, you
can seriously damage the sensors.


OPTICAL PROBES–6026 AND 6136 TURBIDITY; 6025 CHLOROPHYLL; 6131 BGA-PC; 6132 BGA-
PE; 6130 RHODAMINE WT

The 6026, 6136, 6025, 6131, 6132, and 6130 probes require only minimal maintenance. After each

deployment, the optical surface on the tip of the turbidity probe

should be inspected for fouling and cleaned if necessary by

gently wiping the probe face with moist lens cleaning paper. In

addition, for the 6025, 6026, 6136, and 6130 probes, we

recommended replacing the wiper periodically. The frequency

of this replacement depends on the quality of water under

examination. A replacement wiper is supplied with the probes,

along with the small hex driver required for its removal and

reinstallation. Follow the instructions supplied with the probe to

ensure proper installation of the new wiper assembly.

Additional wipers are available from YSI.

PLIERS (SLIP-JAW)

AMMONIUM OR

NITRATE PROBE

NON-WIPING

TURBIDITY PROBE

WIPER

ASSEMBLY

WIPING

OPTICAL PROBE

Advertising
This manual is related to the following products: