plug. Leave the conductivity/temperature and the dissolved oxygen probes in the sonde with a membrane

and electrolyte on the DO sensor. Place enough of deionized, distilled, or tap water in the calibration cup

to cover the sensors, insert the sonde into the vessel, and seal with the cap/O-ring to minimize evaporation.

6820V2-1, 6820V2-2, 6920V2-1, 6920V2-2, 6600V2-2, 6600EDS V2-2, and 6600V2-4 -- Leave the

conductivity/temperature and the dissolved oxygen probes in the sonde with a membrane and electrolyte on

the DO sensor. Remove all other probes from the sonde and store according to the instructions found in the

following section on individual sensors. Cover the empty ports with the provided plugs. Place enough of

deionized, distilled, or tap water in the calibration cup to cover the sensors, insert the sonde into the vessel,

and tighten the threaded cup to attain a good seal and minimize evaporation.

600R and 600QS (with Replaceable Reference Electrode Module) -- Instruments of this design were

generally sold after January, 1996 and can be identified by the presence of 4 probes -- temperature,

dissolved oxygen, pH reference, and pH glass (600R) or pH glass/ORP (600QS) in the bulkhead. Remove

the reference module, store it as described below, and plug the open port with the insert that was provided.

Make certain that the dissolved oxygen sensor has an undamaged membrane and electrolyte in place.

Place approximately 300 mL of tap water in the storage vessel, insert the sonde, and seal the vessel with the

cap and O-ring. Do not use deionized or distilled water in this case, as it may damage the pH glass

sensor that must remain in the sonde.

600 (with Combination pH Sensor) -- Instruments of this design were generally sold prior to January,

1996 and can be identified by the presence of only 3 probes (temperature, dissolved oxygen, pH) in the

bulkhead. Be certain that the dissolved oxygen sensor has an undamaged membrane and electrolyte in

place. Fill the provided storage vessel with a solution that is 2 molar (2 M) in potassium chloride (KCl) to

a level that completely covers the dissolved oxygen and pH probes. See the following section for

instructions on preparation of the KCl storage solution. Seal the vessel with the cap and O-ring.

600 OMS V2-1 – Store sonde dry with optical probe left in port. Cover membrane on ROX sensor with

moist sponge.

All Sondes with Batteries – Because batteries can degrade over time and release battery fluid, it is

extremely important to remove the batteries from all 600XLM, 6920V2-1, 6920V2-2, 6600V2-2,

6600EDS V2-2, 6600V2-4, and 600 OMS V2-1(battery version) prior to long term storage. Failure to

remove batteries can result in corrosive damage to the battery terminals if the batteries happen to leak.


2.10.4

PROBE STORAGE

LONG-TERM STORAGE OF PROBES

The following sections provide additional details on the storage of individual sensors associated with

instruments in the 6-Series product line from YSI.

TEMPERATURE

No special precautions are required. Sensors can be stored dry or wet, as long as solutions in contact with

the thermistor probe are not corrosive (for example, chlorine bleach).

CONDUCTIVITY

No special precautions are required. Sensors can be stored dry or wet, as long as solutions in contact with

thermistor probe and conductivity electrodes are not corrosive (for example, chlorine bleach). However, it

is recommended that the sensor be cleaned with the provided brush prior to long term storage.