2 sonde communication problems – Xylem 6-Series Multiparameter User Manual

Page 284

Advertising
background image

Troubleshooting

Section 6

YSI Incorporated

Environmental Monitoring Systems Operations Manual

6-

2

6.2 SONDE COMMUNICATION PROBLEMS


SYMPTOM

POSSIBLE CAUSE

ACTION

REF

Cannot communicate
with sonde

Sonde not powered

Check 12 vdc source

2.3.3

Cable connection is loose

Check both ends of cable; secure
connectors

2.3.4

Damaged connectors

Check pins at both ends; insure
they are straight, dry and clean.

Com port not selected

Change to other com port, other
peripheral on the same port
(Internal mouse). Try other PC,
650 display/logger or dumb
terminal

2.4.3

Scrambled data

Unmatched baud rate between
host and sonde

Match the baud rate

2.4.3

Host is too slow

Use faster computer

Interface cable failure

Check cable for damage. If
necessary, return for service

2.3.4

Internal failure

Return sonde for service

9