Xylem 6-Series Multiparameter User Manual

Page 251

Advertising
background image

Principles of Operation

Section 5

YSI Incorporated

Environmental Monitoring Systems Manual

5-13

change in voltage. Our experience with ISEs indicates that for best accuracy, the isopotential point should

be determined empirically. To do so, the user employs a third calibration point where the voltage of the

lower concentration standard is determined at a temperature at least 10°C different from the first two

calibration points. The slope, offset, and isopotential point drift slowly, and you should recalibrate the

probe periodically.

All ion selective electrodes are subject to the interaction of species with the sensor membrane, which are

similar in nature to the analyte. For example, chloride ion binds in this way to the nitrate membrane and

produces positive nitrate readings even when no nitrate is present in the medium. Fortunately, most fresh

water does not usually contain significant quantities of ions that produce a large interference on the nitrate

reading, such as azide, perchlorate, and nitrite. It usually does contain some chloride and carbonate ions,

but the interference from these ions is relatively small. For example, if the all of the ionic content of water

with a conductivity of 1.2 mS/cm (Sal = 0.6) were due to the presence of sodium chloride, the nitrate

reading would be erroneously high by about 1.6 mg/L. If the conductivity in this sample were all due to

sodium bicarbonate, the sensor output would indicate the presence of only 0.2 mg/L of non-existent nitrate

from the interference.

Even though the interference from chloride is relatively small and thus tolerable at low salinity, the large
quantity of this species in salt or brackish water creates interference so great as to make the sensor
unsuitable for these media.

Despite the potential problems with interference when using ISEs, it is important to remember that almost

all-interfering species produce an artificially high nitrate reading. Thus, if the sonde indicates the presence

of only small quantities of nitrate, it is unlikely that the reading is erroneously low because of interference.

Unusually high nitrate readings (which could be due to interfering ions) should be confirmed by laboratory

analysis after collection of water samples.

Ion selective electrodes have the greatest tendency to exhibit calibration drift over time of all the sensors

available on the sonde. This drift should not be a major problem for sampling studies where the instrument

can be frequently calibrated. However, if a nitrate sensor is used in a longer-term deployment study with

the sonde, the user should be aware that drift is almost certain to occur. The extent of the drift will vary

depending on the age of the probe, the flow rate at the site, and the quality of the water. For all monitoring

studies using ion selective electrodes, the user should acquire a few “grab samples” during the course of the

deployment for analysis in the laboratory by chemical means or with another nitrate sensor which has been

recently calibrated. Remember that the typical accuracy specification for the sensor (+/- 10 % of the

reading or 2 mg/L, whichever is larger) refers to sampling applications where only minimal time has

elapsed between calibration and field use.

CALIBRATION AND EFFECT OF TEMPERATURE

The nitrate sensor should be calibrated using solutions of known nitrate-nitrogen content according to the

procedures detailed in Sections 2.6.1 and 2.9.2. If a two point calibration protocol is used, the temperature

of the standards should be as close as possible to that of the environmental medium to be monitored. The

recommended calibration procedure is one involving three solutions. Two of the solutions should be at

ambient temperature while the third should be at least 10 degrees Celsius different from ambient

temperature. This protocol minimizes the effects of taking readings at temperatures that are significantly

different from ambient laboratory temperatures.

MEASUREMENT AND CALIBRATION PRECAUTIONS

(1) The temperature response of ion selective electrodes is not as predictable as that of pH sensors.

Therefore, be sure to carry out a 3-point calibration the first time you use the probe. This will provide a

default setting for the effect of temperature on your particular sensor. After this initial calibration, you can

Advertising
This manual is related to the following products: