Xylem 6-Series Multiparameter User Manual

Page 247

Advertising
background image

Principles of Operation

Section 5

YSI Incorporated

Environmental Monitoring Systems Manual

5-9

protocol according to the instructions. The calibration will occur automatically at the end of the specified

warm-up time.

If the Percent Saturation method is selected, the sonde is simply placed in a calibration cup that contains a

small quantity of water or a damp sponge. The probe sensor should not be in the water for this calibration
procedure

. The sonde should be left under these conditions for 10-15 minutes to allow temperature and

humidity equilibration to occur. Then input the true barometric pressure into the sonde software and begin the

calibration protocol according to the instructions. The calibration will occur automatically at the end of the

specified warm-up time.

NOTE: Remember that control of the calibration will be manual rather than automatic if the unit is set up

properly for spot sampling applications (“Autosleep” deactivated).

The DO readings of steady state oxygen systems are greatly affected by temperature (approximately 3% per

degree Celsius) due to the effect of temperature on the diffusion of oxygen through the membrane. The Rapid

Pulse system exhibits a greatly reduced effect of temperature (approximately 1% per degree Celsius), but this

factor still must be accounted for if DO readings acquired at temperatures different from that at calibration are

to be accurate. The sonde software automatically carries out this compensation.

In addition, the relationship between the measured partial pressure of oxygen (percent saturation) and the

solubility of oxygen in mg/L is very temperature dependent. For example, air saturated water (DOsat % =

100) contains 9.09 mg/L at 20

o

C, but only 7.65 mg/L at 30

o

C. The sonde software compensates for both

of these temperature-related factors after instrument calibration. The temperature compensation for the

percent saturation reading is empirically derived, while the conversion from percent saturation and

temperature to solubility in mg/L is carried out automatically by the sonde software using formulae

available in Standard Methods for the Examination of Water and Wastewater (ed. 1989). See Appendix D,

Solubility and Pressure/Altitude Tables for dissolved oxygen solubility tables as a function of salinity

and temperature.

FLOW DEPENDENCE

As noted above, oxygen readings acquired using the Rapid Pulse technology are much less affected by

sample flow than steady state probes. However, there is a finite stirring dependence exhibited by the Rapid

Pulse system if measurements are taken when the probe is being pulsed continuously. Our tests indicate

that, under these sampling conditions, observed dissolved oxygen readings can be 2-3 percent lower than

the true readings in very still water. Minimal movement of the water (which occurs during most

environmental measurements) removes this effect.

This small flow dependence of the sensor is greatly reduced in longer term monitoring deployments where

the sampling interval is longer, e.g. 15 minutes. Under these conditions, the sensor is pulsed for only

approximately 40 seconds every 15 minutes, and normal diffusion of oxygen in the medium re-establishes

the oxygen which has been depleted in the previous warm-up/read sequence.

MEASUREMENT AND CALIBRATION PRECAUTIONS

(1) If water-saturated air is used as the calibrating medium, make certain that both the DO reading and the
temperature

have stabilized (10-15 minutes) before starting the calibration sequence. A wet thermistor can

indicate artificially low temperature readings due to evaporation and this situation will result in poor

temperature compensation and inaccurate readings.

(2) For calibrations in water-saturated air, ensure that the calibration cup being used is vented or pressure

released.

(3) If air-saturated water is used as the calibrating medium, make certain that the water is truly saturated by

sparging it for at least 1 hour with an aquarium air-pump and air-stone.

Advertising
This manual is related to the following products: