Xylem 6-Series Multiparameter User Manual

Page 95

Advertising
background image

Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-89

point calibration, but the software will prompt you to place the sonde in the additional solution to complete

the 3-point procedure.

For all rhodamine calibration procedures, be certain that the standard and sensor are thermally equilibrated

prior to proceeding with the calibration.


REESTABILISHING DEFAULT CALIBRATION – “UNCAL” COMMAND

If, for any reason, you want to return to the factory default settings (slope and offset) for any parameter,

follow the instructions below:

Activate any calibration screen (1-point, 2-point, or 3-point) where a numerical entry is requested. In

the example below, a 2-point turbidity calibration was selected.

Instead of entering a numeric value, type the word “uncal” and press Enter.

An “uncal” entry at any calibration prompt will reset both slope and offset for this parameter to the

factory default settings.

NOTE: When using the 650 MDS as the sonde interface device, the “uncal” operation is performed by

holding the Enter key down and then pressing the Esc key. See Section 3 of this manual for more details.



-----Turbidity calibration--
1-1 point
2-2 point
3-3 point

Select option (0 for previous
menu): 2

Enter 1st Turbid NTU:uncal

Advertising
This manual is related to the following products: