Xylem 6-Series Multiparameter User Manual

Page 85

Advertising
background image

Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-79

which characterize the new membrane and they MUST be entered via the Calibrate|Optic X – Dissolved

Oxy|Enter Cal Sheet selection as show above. Once the new constants have been entered in coded form

they will be transferred to the probe and will be visible in uncoded form in Advanced|Cal Constants.

See Section 5.9, Principles of Operation and Appendix M, ROX Optical DO Sensor for more

information.

User Calibration of the ROX Optical DO Sensor

NOTE CAREFULLY: As opposed to the 6562 Rapid Pulse Polarographic DO sensor described above,

there is no difference between the calibration routine for sensors which will be used for sampling or

monitoring applications. Usually the Autosleep RS-232 feature in the Advanced|Setup menu will be

activated for ROX calibrations, but there is no problem if it is not active.

Select the Optic X-Dissolved Oxy option from the Calibrate menu to calibrate the oxygen probe. The

submenu will offer you the option of calibrating in either percent air-saturation (ODOsat %) or oxygen

concentration (ODO mg/L) units. Calibrating either of the choices will automatically calibrate the other.

After selecting the option of choice (ODOsat % in water-saturated air or air-saturated water is normally

recommended), you will be prompted for the next step where you must elect to do either a 1-point or a 2-

point calibration. Except under unusual circumstances as described below, you should select the 1-point

option for the ROX Optical DO calibration.

Calibration Using Percent Air Saturation – 1-Point

NOTE: YSI recommends that you use this method for calibration of your 6150 Dissolved Oxygen Sensor

to obtain the maximum accuracy under normal operating conditions.

Place the sensor either (a) into a calibration cup containing about 1/8 inch of water which is vented by

loosening the threads or (b) into a container of water which is being continuously sparged with an aquarium

pump and air stone. Wait approximately 10 minutes before proceeding to allow the temperature and

oxygen pressure to equilibrate.

Select ODOsat % and then 1-Point to access the DO calibration procedure. Calibration of your Optical

dissolved oxygen sensor in the DO % procedure also results in calibration of the DO mg/L mode and vice

versa.

Enter the current barometric pressure in mm of Hg. (Inches of Hg x 25.4 = mm Hg).

Note:

Laboratory barometer readings are usually “true” (uncorrected) values of air pressure and can be used

“as is” for oxygen calibration. Weather service readings are usually not “true”, i.e., they are corrected to sea

level, and therefore cannot be used until they are “uncorrected”. An approximate formula for this

“uncorrection” (where the BP readings MUST be in mm Hg) is:

True BP = [Corrected BP] – [2.5 * (Local Altitude in feet above sea level/100)]

Press Enter and the current values of all enabled sensors will appear on the screen and change with time as

they stabilize. Observe the readings under ODOsat %. When they show no significant change for

approximately 30 seconds, press Enter. The screen will indicate that the calibration has been accepted and

prompt you to press Enter again to return to the Calibrate menu.

The minor advantages and disadvantages of calibration in air-saturated water versus water-saturated air are

outlined in Appendix M, ROX Optical DO Sensor

Advertising
This manual is related to the following products: