Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-78

membrane installed on the 6150 probe and NOT a function of the probe, i.e., the constants reflect the

characteristics of the sensor membrane and NOT the probe. When a 6150 probe is purchased from YSI, it

already has a sensor membrane installed and the constants of that membrane are transferred automatically

to the sonde PCB when the sensor is run for the first time. After transfer, the constants can be viewed by

accessing the Advanced|Cal Constants menu as shown below – the regression constants are ODO K2-

ODO K4 and the value at zero oxygen is K1 as shown below.

It should also be noted that the ROX sensor is characterized by two additional parameters – ODO Gain and

Tzero. The ODO gain reflects the small change in the calibration which is carried out be the user in a 1-

point calibration as described below. The Tzero value will always be the same as K1 until the user elects

to perform a 2-point calibration as described below at which time it will change slightly to reflect the

adjustment of the zero oxygen value.

You will also note that a different set of constants for your specific sensor membrane are supplied on the

instruction sheet which came with your 6150 probe. These five constants (K1-K4 and C) have a different

appearance than those shown in Advanced|Cal Constants because they are in coded form to allow error-

free reentry by the user in the unlikely event that the regression constants in the probe are lost. The C

constant is a check sum value associated with the values of K1-K4 and will prevent acceptance of the

constants if typographical errors are made in their entry by the user. The coded constants must be entered

in the Calibrate|Optic X – Dissolved Oxy|Enter Cal Sheet as shown below. The constants can also be

viewed at any time by selecting Calibrate|Optic X – Dissolved Oxy|View Cal Sheet.

YSI recommends that your replace your sensor membrane after 1 year of use by purchasing the 6155

Optical DO Membrane Replacement Kit. This kit will be supplied with a new set of coded constants