Sondes with no optical (turbidity, chlorophyll, or rhodamine WT) probes enabled will display the

following menu.

If any optical probe is enabled in the Sensor menu, then the menu will appear as follows.

Choosing 3-Time Constant will display the following the following or a similar menu to reflect your

optical probe selection:

Recommended settings for time constants are turbidity 12, rhodamine WT 12, chlorophyll 12, BGA-PC 24,

BGA-PE 24, ODO 12, and “other” 4. Note that the time constant can be set independently for each optical

probe and that time constant choices are only available for activated sensors. Rhodamine WT would

appear in the display below if it had been activated as a sensor. All “other” sensors use the same time

constant as shown below.


------------Data filter setup----------
1-(*)Enabled
2-( )Wait for filter
3-Time constant=4
4-Threshold=0.001

Select option (0 for previous menu):


------------Data filter setup----------
1-(*)Enabled
2-( )Wait for filter
3-Time constant. . .
4-Threshold. . .

Select option (0 for previous menu):


------------Time constant ----------
1- Turbid=12
2- Chl=12
3- Other=4


Select option (0 for previous menu):