Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-37

Table 6

6600V2-4 Sonde with Standard Long Calibration Cup****

Probe to Calibrate

Upright

Inverted

Conductivity

525ml

150ml

pH/ORP

500ml

150ml

All Optical Sensors

425ml

DO NOT CALIBRATE***

Table 7

600 OMS V2-1 Sonde* *

Probe to Calibrate

Upright

Inverted

Conductivity

375ml

N/A

Turbidity, Chlorophyll, Rhodamine WT 350ml

N/A

Table 8

600R and 600QS Sondes

Probe to Calibrate

Upright

Inverted

Conductivity

350ml

N/A

pH/ORP

120ml

N/A

* See section below for special instructions dealing with calibration of 6136 turbidity sensor.

** See section below for special instructions dealing with calibration of the conductivity sensor for the 600

OMS V2-1.

*** Optical Sensors CANNOT be calibrated with the sonde in the Upside-Down position because of

interference from the meniscus of the calibration standard.

**** An extended length calibration cup is supplied with the 6600V2-4, 6600V2-2, and 6600EDSV2-2 to

facilitate calibration of the 6136 turbidity sensor. This cup requires the use of larger volumes of other

calibration solutions. Users may choose to purchase the shorter calibration cup sleeve for calibration of

sensors other than the 6136 to reduce the volumes of calibrant. The shorter cal cup sleeve is YSI Item

Number 066267 and can be obtained by contacting YSI Technical Support.



CALIBRATION OF THE 6136 TURBIDITY SENSOR

The 6136 can be calibrated using either the calibration cup supplied with the sonde or with an extended length

calibration cup which can be purchased as an option for the 6820V2-1 and 6920V2-1. An extended cup is

supplied as a standard item with the 6820V2-2, 6920V2-2, 6600V2-4, 6600V2-2, and 6600EDSV2-2 sondes.

If you choose to calibrate with the short calibration cup, you also MUST first make certain that the vessel is

equipped with a BLACK bottom. In addition, you should engage only ONE THREAD when screwing the

calibration cup onto the sonde in order to keep the turbidity probe face as far as possible from the calibration

cup bottom to avoid interference. Even with these techniques, there will still be a small interference from the

bottom of the calibration cup that will cause your field turbidity readings to be approximately 0.5 NTU lower

than the actual reading. This small error is usually only evident when the sonde is deployed in very clear water

where the readings might appear as slightly negative values, e.g., a turbidity of 0.1 NTU would appear as –0.4

NTU.

Use of the extended length cup will require the use of significantly more standard solutions for 6820V2-1/
6920V2-1 sondes (additional 180 mL) if calibration is done in the upright position. To minimize
calibration solution volumes for sensors other than turbidity, users may wish to purchase the shorter
calibration cup sleeve which is supplied as standard with the 6820V2-1 and 6920V2-1. The shorter cal cup
sleeve for any 6600 is Item Number 066267, and the shorter cal cup sleeve for any 6820/6920 is Item
Number 069286. These can be ordered from YSI Technical Support.