Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-7

DO MEMBRANE INSTALLATION WITH THE PROBE INSTALLED IN THE SONDE

Secure the sonde in a vertical position using a vise or a clamp and ring stand such that the sensors are

upright. Remove the probe guard from the sonde.

Remove the old DO membrane and clean the probe tip with water and lens cleaning tissue. Make sure to

remove any debris or deposits from the O-ring groove.

Using the dropper bottle of electrolyte supplied, place electrolyte on the DO probe tip until a high meniscus

is formed as shown in Figure 8 below.

Figure 8

Figure 9

Hold the membrane so that all four corners are supported, but do not stretch the membrane laterally.

Position the membrane over the probe, keeping it parallel to the probe face as shown in Figure 9 above.

Using one continuous downward motion, stretch the membrane over the probe face as shown. See Figure

10 below. Do not hesitate to stretch the membrane.

Figure 10

Figure 11

Install a new O-ring by placing one side of the O-ring in the groove and rolling into place across the

membrane and into the groove on the opposite side of the probe face. Avoid touching the probe face with