Xylem 6-Series Multiparameter User Manual

Page 257

Advertising
background image

Principles of Operation

Section 5

YSI Incorporated

Environmental Monitoring Systems Manual

5-19

turbidity standards supplied by YSI, the value of the standard is 100 NTU when used for calibration of

the 6026 sensor, but 126 NTU when used to calibrate the 6136.

In environmental water that contains suspended particles of varying size and density, the readings for the

two probes will also differ after calibration with either formazin or polymer beads, with the 6026 almost

always showing higher readings and the extent of the difference generally being proportional to the overall

turbidity. For example, if both probes are calibrated at 0 and 100 NTU and then placed in a turbid river, the

6026 might read 400 NTU while the 6136 would read 300 NTU. The absolute difference in the sensor

readings would shrink as the turbidity dropped. Thus, if the 6026 read 30 NTU, the 6136 would likely read

approximately 21 NTU.

Note that the sonde software labels the output of the two sensors with slightly different units of turbidity

(“turbid NTU” for the 6026 and “turbid+ NTU” for the 6136) so that the user will readily be able to

determine which sensor was used in a particular study during later data analysis.

No matter whether the 6026 or 6136 is installed in your sonde, it is important to remember that field optical

measurements are particularly susceptible to fouling, not only from long term build up of biological and

chemical debris, but also to shorter term formation of bubbles from outgassing of the environmental water.

These bubbles can generally be removed in short term sampling applications by simply agitating the sonde

manually. However, for studies longer than a few hours where the user is not present at the site, the quality

of the turbidity data obtained with a turbidity sensor that has no capability of mechanical cleaning is likely

to be poor. However, as noted above, both the 6026 and the 6136 probes are equipped with a mechanical

wiper that makes them ideal for unattended applications. The wiper can be activated in real-time during

discrete sampling operations or will function automatically during long term unattended sampling studies.

The number of wiper movements and the frequency of the cleaning cycle for the unattended mode can be

set in the sonde software. Generally one movement is sufficient for most environmental applications, but

in media with particularly heavy fouling, additional cleaning cycles may be necessary.


CALIBRATION AND EFFECT OF TEMPERATURE

The sonde software offers the option of 1-point, 2-point, or 3-point calibrations procedures. For most

applications, a 2-point calibration at 0 and approximately 100 NTU is sufficient for either the 6026 or the

6136 sensor. However, a user might wish to carry out a 3-point calibration at values of approximately 0,

100 and 1000 NTU to provide maximum accuracy over the entire normally encountered environmental

turbidity range (0-1000 NTU). If the range of turbidity in the environmental sample is well known,

standards of other turbidity values can be utilized (in either 3 point or 2 point routines). However, in all

calibration procedures, one of the standards must be 0 NTU and this should be the first calibration point.

NOTE: Before calibrating your 6026 or 6136 turbidity sensor, pay particular attention to the following

cautions:

For all calibration procedures, you MUST use standards that are based on either formazin or AMCO-

AEPA styrene divinylbenzene beads as described in Standard Methods for the Examination of Water
and Wastewater

and have been prepared either by Hach (formazin based) or AMCO-AEPA based

standards prepared by YSI or an approved YSI vendor listed on the YSI website (www.ysi.com). THE

USE OF STANDARDS FROM OTHER VENDORS AND/OR THOSE PREPARDED FROM

MATERIALS OTHER THAN FORMAZIN OR AMCO-AEPA POLYMER BEADS WILL RESULT

IN BOTH CALIBRATION ERRORS AND INACCURATE FIELD READINGS.

For AMCO-AEPA standards, the value entered by the user during the calibration protocol is

DIFFERENT depending on which sensor (6026 or 6136) is being calibrated. This reflects the

empirically determined fact that the 6026 and 6136 sensors which have been calibrated to the same

value in the primary standard formazin will have different responses in the suspensions of the AEPA-

AMCO beads. This effect is likely due to the larger optical cell volume of the 6136. Thus, for

Advertising
This manual is related to the following products: