Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-15

INSTALLING THE TURBIDITY, CHLOROPHYLL, RHODAMINE WT , BGA-PHYCOCYANIN, BGA-
PHYCOERYTHRIN, AND ROX OPTICAL DISSOLVED OXYGEN PROBES

If you are using any of optical probes listed, it is recommended that the optical sensors be installed first. If

you are not installing one of these probes, do not remove the port plug, and go on to the next probe

installation.

Figure 26

All optical probes, 6136 turbidity, 6025 chlorophyll, 6130

Rhodamine WT, 6131 Phycocyanin Blue-green algae, 6132

Phycoerythrin-Blue-green algae, and 6150 ROX Optical

DO are installed in the same way. Install the probe into the

center port, seating the pins of the two connectors before

you begin to tighten. Tighten the probe nut to the bulkhead

using the short extended end of the tool supplied with the

probe. Do not over-tighten.


CAUTION

: Be careful not to cross-thread the probe nut.

The YSI 6820V2-1 and 6920V2-1 sondes can accept a single turbidity, chlorophyll, Rhodamine WT, BGA-

PC, BGA-PE, or ROX DO probe. The 6600V2-2, 6600EDS V2-2, 6820V2-2, and 6920V2-2 sondes can

accept and utilize two of the six optical sensors at the same time. The two optical ports of these sondes are

labeled “T” and “C” on the sonde bulkhead. Each port can accept any of the six sensors so be sure to

remember which sensor was installed in which port so that you will later be able to set up the sonde

software correctly. The 6600V2-4 sonde can accept and utilize four of the six optical sensors at the same

time. The four optical ports of this sonde are labeled “T” , “C”, “B”, and “O” on the sonde bulkhead. Each

port can accept any of the six sensors so be sure to remember which sensor was installed in which port so

that you will later be able to set up the sonde software correctly.

INSTALLING THE 6562 RAPID PULSE DISSOLVED OXYGEN PROBE, CONDUCTIVITY/TEMP AND
pH/ORP PROBES

Figure 27

Insert the probe into the correct port and gently rotate the

probe until the two connectors align.

The probes have slip nuts that require a small probe

installation tool to tighten the probe. With the connectors

aligned, screw down the probe nut using the long extended

end of the probe installation tool. Do not over-tighten.


CAUTION

: Do not cross thread the probe nut.

OPTIC

PROBE

INSTALLATION

TOOL

PROBE INSTALLATION

TOOL

DO PROBE