Xylem 6-Series Multiparameter User Manual

Page 53

Advertising
background image

Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-47

Before making any field readings, carefully read Section 5.16, Principles of Operation to better

understand the practical aspects of BGA-PC fluorescence measurements.

To begin the calibration, place the correct amount (see Tables 1-8 above) of clear deionized or distilled

water into the YSI clear calibration cup provided. Immerse the sonde in the water. Input the value 0

cells/mL at the prompt, and press Enter. The screen will display real-time readings that will allow you to

determine when the readings have stabilized. Activate the wiper 1-2 times by pressing 3-Clean Optics as

shown on the screen to remove any bubbles from the sensor. After stabilization is complete, press Enter to

“confirm” the calibration and then, as instructed, press Enter to return to the Calibrate menu. Note that

because the range of the BGA sensors in cells/mL is large, the readings may appear to be somewhat noisier

during the calibration procedure than for other 6-series sensors. Variations of +/- 400 cells/mL for the zero

point can be observed with a properly functioning sensor since this value is only 0.2% of the range.

Thoroughly rinse and dry the calibration cups for future use. For additional information related to

calibrating the BGA-PC sensor, see Section 5.16, Principles of Operation.

Note that the 1-point calibration of the BGA-PC sensor in cells/mL will also zero the PC RFU (Relative

Fluorescence Units) parameter which is in units of percent of the full scale of the sensor. Users may wish

to activate PC RFU in the Report menu and simply use this parameter to determine BGA-PC events until a

good correlation between the observed cells/mL value and the value determined from laboratory analysis

has been established.


BGA-PE 1-POINT

Select Optic X BGA-PE - from the Calibrate Menu, and then 1-1 point.

NOTE: This procedure will zero your fluorescence sensor and use the default sensitivity for calculation of

phycoerythrin-containing BGA in cells/mL, allowing quick and easy fluorescence measurements that are

only semi-quantitative with regard to BGA-PE. However, the readings will reflect changes in BGA-PE

from site to site, or over time at a single site.

To increase the accuracy of your BGA-PE measurements, follow the 2-point calibration protocols outlined

in Section 2.9, Sonde Menu.

Before making any field readings, carefully read Section 5.17, Principles of Operation to better

understand the practical aspects of BGA-PE fluorescence measurements.

To begin the calibration, place the correct amount (see Tables 1-8 above) of clear deionized or distilled

water into the YSI clear calibration cup provided. Immerse the sonde in the water. Input the value 0

cells/mL at the prompt, and press Enter. The screen will display real-time readings that will allow you to

determine when the readings have stabilized. Activate the wiper 1-2 times by pressing 3-Clean Optics as

shown on the screen to remove any bubbles from the sensor. After stabilization is complete, press Enter to

“confirm” the calibration and then, as instructed, press Enter to return to the Calibrate menu. Note that

because the range of the BGA sensors in cells/mL is large, the readings may appear to be somewhat noisier

during the calibration procedure than for other 6-series sensors. Variations of +/- 400 cells/mL for the zero

point can be observed with a properly functioning sensor since this value is only 0.2% of the range.

Thoroughly rinse and dry the calibration cups for future use. For additional information related to

calibrating the BGA-PE sensor, see Section 5.17, Principles of Operation.

Note that the 1-point calibration of the BGA-PE sensor in cells/mL will also zero the PE RFU (Relative

Fluorescence Units) parameter which is in units of percent of the full scale of the sensor. Users may wish

to activate PC RFU in the Report menu and simply use this parameter to determine BGA-PE events until a

good correlation between the observed cells/mL value and the value determined from laboratory analysis

has been established.

Advertising
This manual is related to the following products: