Nconfigure hold, Lsimulate – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 100

Advertising
background image

<7. COMMISSIONING OF SC (Conductivity)>

7-8

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

mA

mA

mA

R(1)

R(2)

R(1)

F0503_2.ai

Return mA to Sensor1?

Yes
No

Reset redundant system

Sensor 1

Display

Sensor 2

Alive

Alive

Alive

Alive

Alive

Dead

Dead
Alive

Sensor 1 is dead.

(automatic)

Repair sensor

Redundant system

restart (manual)

If sensor 1 fails, the

output is automatically

switched to the sensor-2

value.

This display is the

example when

“Redundant” is

selected as a

process parameter.

Even if a fault on the sensor 1 is

recovered automatically after

fault detection, the output will not

be switched back automatically

and the sensor 2 value will

continue to be output.

After repairing sensor 1, reset the backup made

by the redundant system. This enables the

sensor 1 value to be incorporated in the output.

On the Reset redundant system, selection of

“Yes” makes the output return to the output of

the 1st module.

Output setup

mA

Output






Configure Hold
Reset redundant

Figure 7.2

Redundant system

Setup
Select one of the output methods: Linear and Table.

Linear:

Set the 0% and 100% values.

Table:

This allows the configuration of an output curve by 21 points (5% intervals).

(The 0% and 100% values must be entered.)

Burn
Select the designated output in case of a fault from among Off, Low, and High. See “7.3 Error

configuration” to set the output.

Off: Output depends on the measured value.
Low: Output is fixed to 3.6 mA
High: Output is fixed to 22.0 mA.

Damping time
This is the time taken for a response to a step input change to reach 90% of the final value

(attenuation time). Set this time in sec.

l

Simulate

When this function is selected, an output of the instrument will be a fixed current value set in % of

the output span. The output span range is -2.5% to 112.5% (3.6 mA to 22.0 mA).
When “Simulate” is selected, regardless of hold setting, the output is always simulated value.

n

Configure Hold

On the Configure Hold, settings are performed to hold of the mA output at a preset value. (Refer

to the section 8.7.) This is enabled only if “mA” is “Output.”

Advertising
This manual is related to the following products: