<4. COMMISSIONING OF pH/ORP>

4-8

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

4.2.5

Impedance settings

This screen is used to set the impedance relating to an input impedance check.
Input 1 impedance represents the “glass membrane impedance” of a pH sensor. In case of an

ORP sensor, it represents “metal electrode impedance.” Input 2 impedance stands for “reference

impedance.”
If you select “Input impedance: Low,” the display moves to the Input impedance screen, enabling

you to set the high and low limits.
The FLXA21 has an impedance check, which is capable of monitoring the impedance of various

sensor systems. In order to “fine tune” this diagnostic tool, it is necessary to set it up to match the

sensors used.
The system can be set to measure the impedances of glass (input 1 impedance: high) and

reference (input 2 impedance: low) electrodes. In applications that tend to leave deposits on the

electrodes or to clog the reference sensor junction, the impedance check (set error configuration)

on the reference sensor can be used to initiate an alarm, or to initiate the wash cleaning process,

if one of the limits is exceeded. Sensors with a liquid earth use this liquid earth in impedance

measurement. Sensors with a liquid earth that Yokogawa can provide include PH8EFP, PH8ERP,

PH8EHP, OR8ERG, OR8EFG, etc. In pure water measurement using PH8EHP, the impedance

of the solution to be measured is high; settings on the Error configuration screen should be set to

Off so that the “Impedance 2 too high” error is not generated.
A sensor without a liquid earth cannot measure impedance. All settings on the Error configuration

screen should be set to Off so that no “Impedance 1 too high” or “Impedance 2 too high” error is

generated.
Sensors not equipped with a liquid earth that Yokogawa can provide include HA405, HA406,

DPA406, DPA405, DPAS405, HF405, HA485, DPA485, etc.
This liquid earth refers to a metal pole built into a sensor or connected externally that contacts the

solution to be measured and transmits the potential of this solution to the converter.

4.2.6 Concentration

Generally pH values are not converted into concentrations. Following linear conversion is not

suitable for a lot of measurements.

The FLXA21 is capable of converting a measured pH value into concentration for display.

Parameters required for concentration calculation can be arbitrarily set by the user. Those

parameters are as follows:

Unit:

The default value is %. The unit can be selected from among mg/L, g/L, and

ppm.

pH:

The default values are 0 and 14 pH (0 to 14). They are the pH zero and span

required for conversion to concentration.

Concentration: The default value are 0 and 100%. Concentration for each pH can be set.

Concentration can be assigned to the pH value zero and span individually.