<5. CALIBRATION OF pH/ORP>

5-3

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

pH

3cal

= ITP -

(273.15 + t

2

)(ITP - pH

2

) + (1 - ) x (273.15 + t

1

)(ITP - pH

1

)

(273.15 + t

3

)

t

3

- t

1

t

2

- t

1

(

)

t

3

- t

1

t

2

- t

1

pH

n

: pH value of n-th buffer solution

T

n

:

Temperature of n-th buffer solution (ºC)

ITP:

ITP value displayed in calibration settings (see section 4.2.4)

Assign 7.00 as the ITP value when a specific value is not available or for the first

calibration of a sensor.

Do not use the 3rd solution whose pH value is within pH

3cal

± 1.

l

zero/slope1,2(3point)

This calibration type is the line-segment type three-point calibration.
If the relation between electromotive force and pH is not in proportion for a wide range, divide the

relevant range into two sections and obtain the zero (asymmetry) and slope (sensitivity) in each

section to perform calibration.

Limitations

• Three different buffer solutions whose difference in pH value between buffer solutions is 1

pH or more should be used.

(1st buffer < 2nd buffer < 3rd buffer or 1st buffer > 2nd buffer > 3rd buffer)
• The temperature difference between the 1st and 2nd buffer solutions should be 20ºC or

less.

• The temperature difference between the 2nd and 3rd buffer solutions should be 20ºC or

less.

5.1.2

Automatic calibration

Calibration can easily be performed by following the calibration menus.
Pre-select the buffer solution to be used from among NIST/DIN 19266, DIN 19267, US, and User

defined buffer in Commissioning → Measurement setup → Calibration settings → pH settings →

Buffers (select set). See also Appendix table 1.
If you select User defined buffer, calibration is performed based on the conditions registered

in buffer tables 1 to 3. Use of the proper buffer table allows the system to perform reliable

calibration.
In the same way as manual pH calibration, select the calibration type from among [zero/slope],

[zero/slope/ITP(3point)], and [zero/slope1,2(3point)].

Calibration is performed stepwise; follow the prompts displayed on the screen.
A stability check is conducted at each measurement point. Proceed to the next step only after the

reading has stabilized.

NOTE

When a sensor or an electrode is exchanged or replaced, sensor wellness data should be reset.
When a sensor is replaced, the replacement can be recorded manually into a logbook. (Refer to

the figure 3.9.)