<7. COMMISSIONING OF SC (Conductivity)>

7-5

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

NOTE

The temperature compensation is not performed around zero. In this case, a warning may be

issued.

7.1.5

Calibration settings

The screen flow differs depending on the combination of objects to be measured.

l

Air adjust limit

Generally, air calibration is not required. To avoid the effects of the cable on the measurement of

lower conductivities such as pure water, a “zero” calibration with a dry sensor may be done.
When using a 4-electrode sensor, additional connections are required. Temporarily interconnect

terminals 13 & 14 with each other and 15 & 16 with each other before making the adjustment.

This is necessary to eliminate the capacitive influence of the cables. The links should be removed

after this step is completed.
As the calibration is performed in air, the resistivity is infinite (open connection). Higher

conductivity values than the air adjust limit indicate that the cell is not in air or is still wet. To

prevent wrong air calibrations, a limit must be given here.

l

c.c. high

High limit of the cell constant expressed as a % of the nominal value. During calibration this value

is used to check if the calibrated cell constant remains within reasonable limits.

l

c.c. low

Low limit of the cell constant expressed as a % of the nominal value. During calibration this value

is used to check if the calibrated cell constant remains within reasonable limits.

l

Step Range

Set the range over which the stability of a measured value is checked. If variations of a measured

value over the stabilization time are within this set value, the measured value is judged to have

stabilized.

l

Stabilization time

During calibration, the stability of the measurement is constantly monitored. When variations of

the value are within a value set in Step Range over this stabilization time set here, the value is

regarded as being stable. If the value does not stabilize within 10 minutes, calibration is aborted.

l

Calibration interval

Set the interval in which a new calibration must take place. If the interval set here is exceeded,

the instrument will be notified according to the setting in “Calib. time exceeded” in the error

configuration.

7.1.6 Concentration

The “Concentration” can be set only when “Conduct. + Concentr.” or “Concentration” is selected

for the Measurement in the Measurement setup menu.
The corresponding concentration given by a measured conductivity can be shown on the Main

display or the Home display.

SC

7