Ndate/time, Nmeasurement setup, Isc 9 – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 115

Advertising
background image

<9. OPERATION OF ISC (Induvtive Conductivity)>

9-3

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

Temp.element Pt1000

Unit

°C

Next

Start quick setup?

Yes
No

Chanage language

Quick setup

Measure Conductivity

Next

Measurement setup

Format YYYY/MM/DD

Date 2010/03/03

Time 17:04:07

Date/Time

Process parameter
Conduct1-TC1

0% value 0.000 nS/cm
100% value 1.000 S/cm

Finish

mA(output)

Temperature settings

Measuring unit /cm
Cell constant 1.880 /cm
(factory)

Next

Configure sensor

°C
°F

YYYY/MM/DD
MM/DD/YYYY
DD/MM/YYYY

Conductivity
Concentration
Conduct.+Concenter.

Conduct1-TC1
Temperature1
Concent1-TC1
Conduct1-TC2
Concent1-TC2

/cm
/m

Next

Next

Next

Finish

Next

Monitor display

Pt1000
NTC30k

Figure 9.2

Quick setup

n

Date/Time

The date display format can be selected from among the three types.
Enter the date or time of day by using the numerical keys.
For details, see section 10.5.4.

n

Measurement setup

Select a suitable measurement parameter from among those displayed and set it up.
Measurement parameter setup can be made only when “ISC” is selected on the Sensor setup

screen.
For details, see section 10.1.1.

ISC

9