<17. CALIBRATION OF SENCOM pH/ORP>

17-6

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

17.4 HOLD

The FLXA21 has a function to hold the mA output at a preset value (default: “Last”).
Use this menu to hold the output.

For the settings, see “n Configure Hold” on page 16-11.
During commissioning or quick setup, the output is automatically held.
Setting “Hold during Calibration/Wash” to “Disabled” deactivates the hold function during

calibration or washing.

Press

to select Execute: HOLD and then choose Manual Hold ON or Manual Hold OFF.

This allows you to set up manual hold.

10.38

Tag:SENCOM

25.0

4mA

20mA

PH1

°C

pH

HOLD

Figure 17.2

Example of the display with the manual hold enabled

To cancel manual hold, press the lit

HOLD

section on the Main display.