6 temperature calibration, 7 hold, Temperature calibration -5 – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 140: 7 hold -5, Isc 11

Advertising
background image

<11. CALIBRATION OF ISC (Inductive Conductivity)>

11-5

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

11.6 Temperature calibration

In order to make the most accurate measurements, it is important to have a precise temperature

measurement. This affects the display of temperature, and the output signal (when used). More

important, however, is the temperature compensation, and calibration accuracy. The temperature

of the sensor system should be measured independently with a high precision thermometer.

The value of “Measured temperature” should then be adjusted to agree with the reading (zero

offset calibration only). For best accuracy, this should be done as near to the normal operating

temperature as possible.

11.7 HOLD

The FLXA21 has a function to hold the mA output at a preset value (default: “Last”).
Use this menu to hold the output.

For the settings, see “n Configure Hold” on page 10-7.
During commissioning or quick setup, the output is automatically held.
Setting “Hold during Calibration” to “Disabled” deactivates the hold function during calibration.

Press

to select Execute: HOLD and then choose Manual Hold ON or Manual Hold OFF.

This allows you to set up manual hold.

10.00

Tag:FLXA21–ISC

25.0

4mA

20mA

Conduct1-TC1

°C

mS/cm

HOLD

Figure 11.10

Example of the display with the manual hold enabled

To cancel manual hold, press the lit

HOLD

section on the Main display.