<7. COMMISSIONING OF SC (Conductivity)>

7-13

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

For more information on HART communication, see the Technical Information (TI 12A01A02-

60E) in the attached CD-ROM.

n

PH201G

Select this menu if the PH201G distributor is connected to the instrument.
In the PH201G setup screen, make settings for “Hold contact,” and “Fail contact.”

l

Hold contact

Select Disabled or Enabled.
When this item is enabled, the output will be held according to the setting of “Hold type” on the

Hold setup screen.

l

Fail contact

Select a status from among “Fail + Warn,” “Fail only,” and “Disabled.”
This setting depends on the error configuration. See “7.3 Error configuration.”
“Fail” corresponds to “Fault.”

7.5.6

Factory setup

For “Factory setup,” there is no item to be set by the user.

NOTE

This menu is for service engineers only. This section is protected by a password. Attempting to

change data in the factory adjustment menu without the proper instructions and equipment could

result in corruption of the instrument setup and damage the performance of the unit.

7.6

Display setup

This screen is used to make various settings relating to screen display.

NOTE

Settable items differ depending on settings in “Measurement setup.”

7.6.1

Main display (Dual display, Individual display)

Main display
Trend
Auto Return
Ad.just contrast
MONITOR display

Display setup

Dual display
Individual display
Trend
Auto Return
Adjust contrast
MONITOR display

or

Display setup

Figure 7.3

Display setup screens for a single module (left) and two modules (right)