4 calibration settings, Calibration settings -6, Nph settings – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 189

Advertising
background image

<16. COMMISSIONING OF SENCOM pH/ORP>

16-6

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

l

NEN6411

This algorithm takes into account the dissociation of water in strong acid and strong alkaline

solutions. It is particularly useful for pH measurement of boiler feed water.

16.2.4

Calibration settings

The screen flow differs depending on the combination of objects to be measured.
Calibration settings for a pH converter involve slope (sensitivity), zero (aspot), and ITP

(isothermal point). Figure 16.2 shows the pH value against the mV output of the sensor. The

characteristic for pH measurement is an offset also known as aspot [mV] or zero [pH] and a

Slope [%, mV/pH]. For an ideal sensor, the theoretical slope is 59.16 mV/pH at 25ºC. The slope

can be entered in mV/pH or as a percentage of the theoretical slope (100% corresponds to

59.16 mV/pH). ITP represents a pH value where the output of the sensor does not change with

temperature. Note that slope and zero are defined at 25ºC.

ΔpH

ΔmV

ITP (Isothermal point)

Aspot

0 mV

0 °C

10 °C

25 °C

mV

500

0

- 200

14

pH

ITP Zero

0

7

Figure 16.2

Calibration parameters

n

pH settings

l

Zero and Slope units

Zero is an alternative to Asymmetry Potential. The method of zero (aspot) unit conforms to the

DIN standard for IEC 60746-2 instruments. Zero is defined in pH or mV.
The unit of Slope (sensitivity) is mV/pH or % (with the theoretical value as 100%).

l

Limits and timing

Zero High/Low
Zero (aspot) high and low limits. During calibration, it is checked whether the new zero exceeds

these high and low limits. Narrowing the band will prevent bad calibration procedures and

calibration of bad sensors, resulting in higher accuracy. The default values should be adjusted to

suit the application and the “users” criterion.
Slope High/Low
Slope (sensitivity) high and low limits. During calibration, it is checked whether the new slope

exceeds these high and low limits. Narrowing the band will prevent bad calibration procedures

and calibration of bad sensors, resulting in higher accuracy. The default values should be

adjusted to suit the application and the “users” criterion.

Advertising
This manual is related to the following products: