<Appendix 2 For SC (Conductivity)>

App.2-12

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

The USP<645> limit value is a non-temperature-compensated conductivity. When a measuring

conductivity needs to be shown on the main or home display as a non-temperature-compensated

conductivity, set the temperature compensation on “None”. (Refer to the section 7.1.4.)

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

25

50

75

100

USP safety margin

(user programmable)

USP limit

Temperature (°C)

Conductivity (µS/cm)

Figure 3 Conductivity limit value and temperature directed by USP<645>

n

Changing the settings

If any setting is accidentally changed, values to the right of the relevant arrow in Table 6 are all

initialized.

Table 6 Parameters that initialize other values

Measurement ->

Output: Process

parameter ->

Linear: 0% value, 100% value
Table
Communication: HART: PV

Display setup: Individual display (Main display)
Trend Graph Screen ->

Y-axis (low, high)

Communication: HART

Configure sensor:

Measuuring unit ->

Display setup: Individual display (Main display): unit