<16. COMMISSIONING OF SENCOM pH/ORP>

16-8

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

Time

Calibration

Zero, Slope

a

b

c

d

a: Actual change

b: Zero and slope calculated from

the calibration data

c: Zero and slope calculated from

the calibration data and shifted

using the latest calibration data

d: Data from calibration

Figure 16.3

Auto correct

n

ORP settings

l

Limits and timing

Zero High/Low
Zero (aspot) high and low limits. During calibration, it is checked whether the new zero exceeds

these high and low limits. Narrowing the band will prevent bad calibration procedures and

calibration of bad sensors, resulting in higher accuracy. The default values should be adjusted to

suit the application and the “users” criterion.

Slope High/Low
Slope (sensitivity) high and low limits. During calibration, it is checked whether the new slope

exceeds these high and low limits. Narrowing the band will prevent bad calibration procedures

and calibration of bad sensors, resulting in higher accuracy. The default values should be

adjusted to suit the application and the “users” criterion.

Step Range
Set the range over which the stability of a measured value is checked. If variations of a measured

value over the stabilization time are within this setpoint, the measured value is judged to have

stabilized.

Stabilization time
This is the time over which the stability of an ORP value is monitored during calibration. When

variations of the ORP value are within the value set in Step Range over this stabilization time

set here, the value is regarded as being stable. If the value has not stabilized within 10 minutes,

calibration is aborted.

Calibr. interval
Set the interval in which a new calibration must take place. If the interval set here is exceeded,

the instrument will give a warning or fault according to the setting on the Error configuration

screen.

l

Zero/Slope

You can enter zero and slope values directly.